Procedure :
1. Put 1 ml of the glucose reagent in each of two small test tubes.
2. Withdraw 10? of serum by micropipette and put them in one of the two  tubes and sign it as test tube.
3. Withdraw 10? of glucose standard and put them in the other tube and  sign it as standard tube.
4. Leave the two tubes in water bath for 10 minutes or in room  temperature for 20 minutes.
5. Use the wave length 500 nm for measuring.
6. Put the standard and record the reading then put the tested serum and  record  the reading .
Calculations:
[The reading of test  / The reading of standard ] * conc. of glucose  standard .
(usually 100mg/dl)
Normal values :
FBS        (70 -115  mg/dl)      fast blood sugar (fasting 8 hrs) .
RBS        (up to 125 mg/dl)    random blood sugar
PPBS      (80 – 140 mg/dl)     postprandial blood sugar (2 hrs  after meal)